Famos - Bowie Dick Test Set for both Hollow & Porous Loads - Hospital Sterilisers and +60 litre chamber

Bowie Dick Test - Famos - Helix TypeThe Famos Brand BD Test Kit is a complete Bowie-Dick Simulation Test Kit that complies with and is validated according to EN 285 (test report available on request).

The European Community Standard - EN 867 and the relevant Australian/New Zealand Standards AS/NZS 4187 that a Bowie-Dick Test shall be performed on B Class types of steam sterilizers (Pre- Vacuum) equiped with a process recorder/printer "On Commissioning, at Installation; at calibration; after repair; Daily."

The Bowie-Dick Test should be performed prior to the first batch of the day, and the responsible operator is to record the results in the Sterilizer Log System. If the test proves negative the sterilizer is to be shut down and "corrective action taken."

Advantages of The Famos Bowie-Dick Simulation Test Set

Validated to test pack DIN EN 285 Tests air removal. Tests the efficiency of the sterilizer's air removal system. Tests for the presence of non-condensable gasses in the steam (eg air, carbon dioxide)

  • Detects possible leaks in the sterilizer system.
    Tests steam quality (superheated, wet steam).
    Monitors Temperature and Time
  • Superior performance compared to porous test systems.
    The Famos B-D Simulation Test Kit proves:
    - Steam penetration into porous loads
    - Steam penetration into Tubes eg drills, hand pieces, instrument hollows (lumen), suction devices etc.
  • EN 285 validation test results available.
  • May be used in Class B sterilizers that do not incorporate a Bowie-Dick Test Cycle
  • Economical with a reusable test tube (1000 cycle life span) and less
    expensive indicator strips.

Daily Air Removal/Steam Penetration

The efficiency and removall of air is a critical parameter for the proper function of pre-vacuum (Class B) sterilisers. A steam penetration test should be performed in an cmpty steriliser at the beginning of every day to document the efficiency of the vacuum pumps and to indicate possible leaks and/or poor steam quality.

  • Get a more sensitive warning of possible air leaks with 6 test pads to indicate pass or failure
  • Brass test tube gives a component life of 1000 cycles (with correct replacement of wear parts).
  • Validation according to test pack DIM EM 285.
  • Colour Control Indicator has adhesive tape for attaching directly to B-D test record system. eg The Trakka Sterilizer Record & Tracking System.


order information:

CODE DESCRIPTION
15.100.072 Famos Bowie Dick Test Starter Pack (+60 litre chamber)
  Stainless Steel Helix Device, 100 Chemical Indicators, 1 x Cotton Bag, Instructions to use
15.100.050 Bowie Dick Refill Pack Refill for 15.100.072 (+60 litre chamber)
  250 Chemical Indicators, 1 x Washer. 1 x Silcone Junction ,
  1 x Cotton Bag, Instructions to use